Chapter 2: Description
This chapter contains the description of the NomoLine ISA CO2 physical features.
General System Description
The NomoLine ISA CO2 system includes the following:
NomoLine ISA CO2 Device
NomoLine Sampling Line
For a complete list of compatible NomoLine sampling lines, cannulas, and related
consumables, visit www.masimo.com.
Features
Front View
1. NomoLine Capnography Input Connector
Allows connection to NomoLine® for capnography
measurements. Light Emitting Gas Inlet (LEGI) Indicator provides
visual indications of capnography status. See LEGI Indicator on
page 19.
Rear View
2. Capnography Gas Sample Exhaust Port
Exhaust port for gas samples. See System Setup on page 17.
www.masimo.com
15
Masimo
NomoLine ISA CO2
NomoLine Sampling Lines
NomoLine ISA CO2 samples gas from the respiratory circuit through the NomoLine Family
sampling line at a rate of 50 sml/min, making measurements of CO
pediatric, infant and neonatal patients. The NomoLine Family of sampling lines are designed
for optimal performance and measurement fidelity when used with the NomoLine ISA gas
analyzers.
NomoLine sampling lines include nasal and nasal/oral cannulas for non-intubated patients
with and without supplementary oxygen delivery and airway adapter sets for intubated
patients.
As long as no sampling line is connected, the NomoLine ISA gas analyzer remains in a
low-power sleep mode. Once the sampling line is connected, the NomoLine ISA gas analyzer
switches to measuring mode and starts delivering gas data.
For ordering information about NomoLine sampling lines, cannulas, and related
consumables, visit www.masimo.com.
Sampling Line Replacement
NomoLine sampling lines should be replaced between each patient or when the sampling
line becomes occluded. Occlusion occurs when water, secretions etc. are aspirated from the
respiratory circuit to such an extent that NomoLine ISA CO2 cannot maintain the normal 50
ml/min sample flow. This is indicated by a red flashing LEGI indicator and an alarm message;
replace the sampling line and wait until the LEGI indicator switches to green, indicating that
the NomoLine ISA CO2 gas analyzer is again ready for use.
For ordering information about NomoLine sampling lines and related consumables, visit
www.masimo.com.
www.masimo.com
Chapter 2: Description
possible for adult,
2
16
Masimo