Note: Use and store the NomoLine ISA CO2 in accordance with specifications. See the
Specifications section in this manual.
Performance Warnings and Cautions
WARNING: NomoLine ISA CO2 should not be used as the sole basis for medical decisions. It
must be used in conjunction with clinical signs and symptoms.
WARNING: Use of high-frequency electrosurgical equipment in the vicinity of NomoLine ISA
CO2 may produce interference and cause incorrect measurements.
WARNING: Do not use the NomoLine ISA CO2 with metered-dose inhalers or nebulized
medications as this may clog the bacteria filter.
WARNING: Properly apply sampling lines according to the sampling lines directions for use.
Misapplied sampling lines that become partially dislodged may cause no or incorrect
readings.
WARNING: Replace the sampling line if the sampling line input connector starts flashing red,
or host device displays a check sampling line type of message.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are operating
properly.
WARNING: Portable and mobile RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30cm (12 inches) to
any part of the NomoLine ISA CO2 including the cable. Otherwise, degradation of the
performance of the NomoLine ISA CO2 could result.
WARNING: Too strong positive or negative pressure in the patient circuit might affect the
sample flow.
WARNING: Strong scavenging suction pressure might affect the sample flow.
CAUTION: Do not operate NomoLine ISA CO2 outside of the specific operating environment.
CAUTION: NomoLine ISA CO2 should be mounted securely to avoid risk of damage to the
NomoLine ISA CO2.
Cleaning and Service Warnings and Cautions
WARNING: To avoid electric shock, always physically disconnect the NomoLine ISA CO2 and
all patient connections before cleaning.
WARNING: Do not attempt to remanufacture, recondition or recycle the NomoLine ISA CO2
as these processes may damage the electrical components, potentially leading to patient
harm.
CAUTION: Do not sterilize or immerse NomoLine Family sampling lines in liquid.
CAUTION: To avoid permanent damage to the NomoLine ISA CO2, do not use undiluted
bleach (5% - 5.25% sodium hypochlorite) or any other cleaning solution not recommended.
www.masimo.com
Safety Information, Warnings and Cautions
10
Masimo
NomoLine ISA CO2
CAUTION: Do not use petroleum-based or acetone solutions, or other harsh solvents, to clean
the NomoLine ISA CO2. These substances affect the device's materials and device failure can
result.
CAUTION: Do not submerge the NomoLine ISA CO2 in any cleaning solution or attempt to
sterilize by autoclave, irradiation, steam, gas, ethylene oxide or any other method. This will
seriously damage the device.
CAUTION: To prevent damage, do not soak or immerse NomoLine ISA CO2 in any liquid
solution.
Compliance Warnings and Cautions
WARNING: Any changes or modifications not expressly approved by Masimo shall void the
warranty for this equipment and could void the user's authority to operate the equipment.
WARNING: Use of accessories and cables other than those specified or provided could result
in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
WARNING: Dispose NomoLine Family sampling lines in accordance with local regulations for
biohazardous waste.
CAUTION: Disposal of Product: Comply with local laws in the disposal of the device and/or its
accessories.
CAUTION: For FCC compliance information, refer to the host monitor Operator's Manual.
Note: Use the NomoLine ISA CO2 in accordance with the Environmental Specifications
section in the host monitor Operator's Manual.
www.masimo.com
Safety Information, Warnings and Cautions
11
Masimo