NomoLine ISA CO2
Equipment Classification per IEC 60601-1
Type of Protection
Class II
Degree of Protection of
Defibrillation proof BF-Applied Part
Electrical Shock
Protection against harm
IP34, Protection from ingress of particulates ≥ 2.5 mm and
from liquid ingress
against splashing water from all directions
Mode of Operation
Continuous operation
Additional Specifications
General
Specifications
Ambient CO
≤ 800 ppm (0.08 vol%)
2
Recovery Time After
Unaffected
Defibrillator Test
Drift of measurement Accuracy
No drift
Interface
RS-232 and USB
Water Handling
NomoLine Family sampling lines with proprietary water
removal tubing.
Sampling Flow Rate*, **
50 ± 10 sml/min
*Volumetric flow rate of air corrected to standardized conditions of temperature and pressure.
** Flow accuracy specification for the extended temperature range (-20°C to 0°C) is +15/-10
sml/min.
Data Output
Specifications
Breath
Adaptive threshold, minimum 1 vol% change in CO
detection
Respiration
0 to 150 ± 1 breaths/min
rate ***
www.masimo.com
Chapter 6: Specifications
concentration.
2
25
Masimo
NomoLine ISA CO2
Data Output
Specifications
Fi and ET **** Fi and ET are displayed after one breath and have a continuously updated
breath average.
The following method is used to calculate end-tidal (ET) values:
The highest concentration of CO
weight function applied to favor values closer to the end of the cycle.
ET will typically decrease below nominal value (ET
(RR) exceeds the RR threshold (RR
CO
:
2
(with NomoLine HH Adult/Pediatric Airway Adapter Set)
Flags
Breath Detected, No Breath Detected, Check Sampling Line, Unspecified
Accuracy, Sensor Error
*** Measured at I/E ratio 1:1 using breath simulator according to EN ISO 80601-2-55 fig.
201.101.
**** Measured according to EN ISO 80601-2-55.
Gas Analyzer
Specifications
Sensor head
Dual channel NDIR type gas analyzer measuring at
3.5 to 4.5 μm.
Data acquisition rate 10 kHz (sample rate 20 Hz /
channel).
Compensations
Automatic compensation for pressure and
temperature.
Manual compensation for broadening effects on CO
Calibration
No span calibration is required.*****
Warm-up time
< 10 seconds (Concentrations reported and full
accuracy)
CO2 Rise time at 50 sml/min sample
≤ 200 ms
flow *****
Analyzer System Response Time
< 3 seconds
***** An automatic zeroing is performed. See Zeroing on page 34.
www.masimo.com
Chapter 6: Specifications
during one breathing cycle with a
2
) when respiration rate
nom
) according to the following formula for
th
26
Masimo
.