NomoLine ISA CO2
Symbol
Description
Medical Device
Class II Equipment
Instructions/Directions for Use/Manuals are available in electronic format
@http://www.Masimo.com/TechDocs
Note: eIFU is not available in all countries.
www.masimo.com
Chapter 6: Specifications
Symbol
Description
Authorized representative in
the European community
China Restriction of
Hazardous Substances
31
Masimo
Chapter 7: Service and Maintenance
Cleaning
Cleaning of the NomoLine ISA CO2 should be performed at regular intervals or in accordance
with hospital, as well as local and governmental regulations.
See Safety Information, Warnings and Cautions on page 9.
WARNING: To avoid electric shock, always physically disconnect the NomoLine ISA CO2 and
all patient connections before cleaning.
CAUTION: To avoid permanent damage to the NomoLine ISA CO2, do not use undiluted
bleach (5% - 5.25% sodium hypochlorite) or any other cleaning solution not recommended.
CAUTION: Do not use petroleum-based or acetone solutions, or other harsh solvents, to clean
the NomoLine ISA CO2. These substances affect the device's materials and device failure can
result.
CAUTION: Do not submerge the NomoLine ISA CO2 in any cleaning solution or attempt to
sterilize by autoclave, irradiation, steam, gas, ethylene oxide or any other method. This will
seriously damage the device.
CAUTION: To prevent damage, do not soak or immerse NomoLine ISA CO2 in any liquid
solution.
Note: To prevent cleaning liquids and dust from entering the NomoLine capnography gas
analyzer through its sampling gas inlet connector, keep the sampling line fitted while
cleaning NomoLine ISA CO2.
NomoLine ISA CO2 may be cleaned using a cloth moistened (not wet) with the cleaning
agent. The surfaces of the NomoLine ISA CO2 may be cleaned with the following solution(s):
70% ethyl alcohol
70% isopropyl alcohol
Glutaraldehyde Solution
Quaternary Ammonium Chloride Wipe
0.5% Sodium Hypochlorite/Water Solution
Accelerated Hydrogen Peroxide
Maintenance
Once every year it is recommended to perform maintenance on NomoLine ISA CO2. A
NomoLine ISA Maintenance Kit containing components and instructions to perform
maintenance procedures is available through www.masimo.com. The NomoLine ISA
Maintenance Kit may also be used for troubleshooting.
Operational Check
In addition to the Pre-use check (see System Setup on page 17), the Operational check
includes a Capnogram check and an Occlusion check. The Operational check shall be
performed according to the following procedure before the NomoLine ISA CO2 is taken into
operation or as part of troubleshooting.
www.masimo.com
33