7.6
Reinitializing the Analyzer
If the analyzer requires reinitialization, use the reinitialization function to return all settings to their
original factory default values (including the clock), and also to clear all patient results, control results,
and custom reference ranges from memory. Perform the following steps to reinitialize the analyzer.
1.
In the Home screen, select the Settings icon.
2.
Select the More Settings icon.
3.
Select the Factory Settings icon.
4.
Archive the analyzer's reference ranges — including custom reference ranges — so they can
be restored after the analyzer is reinitialized. For instructions, see
Reference Ranges" on page
CAUTION:
The next step is irreversible. All settings — including custom reference ranges — are
returned to their factory values, and the memory is cleared. Results previously stored,
however, are not affected.
5.
Select
YES
to clear all memory from the analyzer.
6.
Refer to the following sections to reset any custom reference ranges, the date and time, or the
units of measurement:
"Customizing Reference Ranges" on page 4-2
"Changing Date and Time" on page 4-15
"Selecting Units" on page 4-17
7-4
4-11.
to restore all factory settings (including the clock) and
Troubleshooting
"Printing and Archiving