7.4
Disc Cancellations
If the disc cancels, record the following information or print an error report and contact Abaxis Techni-
cal Support:
Lot number
Product name
4-digit error code
Specimen type and sample
7.5
Instrument and Result Messages
The Piccolo Xpress chemistry analyzer performs a series of internal quality checks to ensure accurate
results. When the analyzer detects a problem, the analyzer will either suppress certain chemistry results
(
,
,
, or
~~~
HEM
LIP
ICT
printed). When either of these situations occur, the operator should proceed as described below.
7.5.1
Chemistry Suppression
In this situation, patient results print, but some results do not have numerical values. Instead, the results
include troubleshooting flags
If this happens, record the error code, print an error report using the Recall function (see
Results" on page
5-1), and contact Abaxis Technical Support to review the error report.
7.5.2
Disc Cancellation
In this situation, the display shows
shows an error code and a brief description of the error.
If this happens, record the error code, print an error report using the Recall function (see
Results" on page
5-1), and contact Abaxis Technical Support to review the error report.
7-2
printed in place of values) or cancel the run (disc cancels and no patient results
,
,
~~~
HEM
LIP
Run Canceled
Troubleshooting
, or
printed in place of values.
ICT
and does not print patient results. The display also
"Recalling
"Recalling