3.2
Preparing the Reagent Disc
See
"Reagent Discs" on page 2-7
including handling instructions. Make sure you are thoroughly familiar with this infor-
mation before beginning to use the analyzer.
See
"Sample Requirements" on page 3-2
CAUTION:
Analysis must begin immediately (no more than 10 minutes) after dispensing the
sample into the reagent disc.
CAUTION:
Discs are fragile - always handle with care. Do not tap discs on the table or work
bench to empty the sample port. Do not use a disk that has been dropped.
CAUTION:
Inspect every reagent disc for damage before use. Never use a damaged disc.
3.2.1
Step 1 — Dispense the Sample
Use a micropipette (one is included with the Piccolo Xpress) or other transfer device to dispense
approximately 100 µL of sample into the disc via the sample chamber.
CAUTION:
Wear powder-free gloves while handling reagent discs or operating the analyzer.
Powder can interfere with the analyzer's optical components.
1.
Fill the sample chamber.
a.
3-4
Using the Piccolo 100 µl volume pipette, firmly attach a new tip to the end of
the pipette.
Testing and Results
for complete information about Piccolo reagent discs,
for sample handling and storage requirements.
Write patient
ID here