3.1
Sample Requirements
The Piccolo Xpress chemistry analyzer accepts lithium-heparinized whole blood,
plasma, or serum samples.
Lithium heparin is the only anticoagulant recommended for use with the Piccolo
Xpress.
Note:
When collecting the sample in lithium heparin collection tubes, fill the tube at least half-
way so the anticoagulant does not become too concentrated in the sample.
If using CLIA '88 Waived reagent discs: only lithium-heparinized whole blood can be
used as a sample.
If performing a Basic Metabolic Panel Plus: only lithium-heparinized plasma or
serum can be used as a sample. Heparinized whole blood cannot be used.
A sample size of 90–120 µL is required.
Whole blood must be analyzed within 60 minutes of collection, or separated into plasma
or serum.
To prevent hemolysis, do not refrigerate or shake whole blood.
Plasma or serum can be stored at room temperature for no longer than 5 hours after cen-
trifugation. If storage for more than 5 hours is required, refrigerate the sample in the
stoppered tube at 2-8 °C (36-46 °F) for no longer than 48 hours, or store it at -10 °C for
up to 5 weeks in a freezer with no self-defrost cycle. Under these conditions, there will
be no clinically important changes in most analyte concentrations.
For accurate interpretation of glucose and lipid results, the patient should fast for at least
12 hours before the sample is collected.
WARNING: O
PERATOR HEALTH AND SAFETY REGULATIONS REQUIRE THAT
TIONS BE OBSERVED AT ALL TIMES WHILE HANDLING HUMAN BLOOD SAMPLES OR
WORKING WITH THE
F
OR DETAILS
S
AFETY AND
H
AZARDOUS
GOING TO
WARNING: O
PERATORS MUST ALSO BE AWARE OF STATE AND LOCAL
LATIONS THAT MAY SPECIFY REQUIREMENTS IN ADDITION TO THE FEDERAL REGU
LATIONS
3-2
P
ICCOLO
,
PLEASE REFER TO
H
S
EALTH
TANDARDS
S
: B
UBSTANCES
://
.
HTTP
WWW
OSHA
,
AND MUST ALSO ENSURE COMPLIANCE PROCEDURES ARE IN PLACE
Testing and Results
X
PRESS CHEMISTRY ANALYZER IN ANY WAY
OSHA 29 CFR P
"), S
TANDARD
P
LOODBORNE
ATHOGENS
.
GOV
AND SEARCHING FOR
U
NIVERSAL
1910 ("O
ART
CCUPATIONAL
N
1910.1030 ("T
UMBER
"),
WHICH CAN BE FOUND BY
"1910.1030".
OSHA
BIOHAZARD REGU
P
-
RECAU
.
OXIC AND
-
-
.