Sensor board replacement

Tools required:
Zipper pull tool or equivalent
Needle nose pliers
ESD system
Procedure:
1.
See
Foot box cover access on page
2.
Properly ground yourself (see
3.
Disconnect all six of the pod sensor hoses from the foot box by gently pushing inward on the coupling and pulling
outward on each of the hoses.
Notes
When handling the pod sensor hoses, do not bend or kink the hoses.
Pay attention to the sensor hose position and insertion color and number labels.
During re-install, make sure to insert the hose and once you hit a stop, continue pushing until you reach a hard
stop.
4.
Disconnect the cable (AN) from the sensor board (AB)
5.
Using needle nose pliers, lifting up slightly, disconnect the sensor board (AB) from the three plastic standoffs
1-30 on page
1-57).
Note: During reinstall, align the standoffs to the main board and press near the standoffs to seat the main board. Do
not bend the board.
6.
Remove and discard the sensor board.
7.
Reverse steps to reinstall.
Notes
During reinstall, make sure to align the Velcro® patches.
After installation, remove the zipper pull tool or equivalent from the zipper.
Make sure that the watershed covers the zipper.
www.stryker.com
Service
1-40.
Protecting against Electrostatic Discharge (ESD) on page
(Figure 1-30 on page
Figure 1-30: Sensor board
2971-109-0 01 REV B
1-33).
1-57).
E E N N
(Figure
1-57