Pod assembly replacement (Continued)
11. Separate the Velcro® from the head end of the pod assembly (C) and feed the turn bladder hose quick connection
through the foam crib.
12. Remove and discard the pod assembly
13. Reverse to install the new pod assembly.
Notes
During reinstall, make sure to align the Velcro® patches.
After installation, remove the zipper pull tool or equivalent from the zipper.
Make sure that the watershed covers the zipper.
14. Run the leak diagnostic test (see Accessing the Isolibrium Diagnostic Menu in the InTouch maintenance manual).
15. Verify proper operation before returning the product to service.
www.stryker.com
Service
Figure 1-15: Pod assembly
(C)(Figure 1-15 on page
1-39).
2971-109-0 01 REV B
E E N N
1-39