Les langues disponibles

Les langues disponibles

Power cord replacement (Continued)
6.
Disconnect the power cord quick connect (F) from the power supply
7.
Feed the two pin connector through the nut (E)
8.
Remove and discard the power cord (AV)
9.
Reverse steps to reinstall.
Notes
During reinstall, first assemble the power cord ground star washers and nut, and then the power supply ground
star washers and nut.
During reinstall, make sure to align the Velcro® patches.
After installation, remove the zipper pull tool or equivalent from the zipper.
Make sure that the watershed covers the zipper.
10. Verify proper operation before returning the product to service.