Main power board assembly (Continued)
E E N N
Cable and test point
(LED) location
SLI (D33)
SLD (D32)
SRI (D25)
SRD (D24)
IN (D28)
EX (D29)
J14
J6
1-30
Circuit boards
Voltage
Positive lead
+24VDC
TP12
+24VDC
TP19
+24VDC
TP8
+24VDC
TP15
+24VDC
TP10
+24VDC
TP17
+12VDC
Pin 1 red
+24VDC
Pin 1 yellow
2971-109-001 REV B
Negative lead
Description
TP24 GND
+24VDC from the
main board to
solenoid valve (Seat
Left Inflate)
TP24 GND
+24VDC from the
main board to
solenoid valve (Seat
Left Deflate)
TP24 GND
+24VDC from the
main board to
solenoid valve (Seat
Right Inflate)
TP24 GND
+24VDC from the
main board to
solenoid valve (Seat
Right Deflate)
TP24 GND
+24VDC from the
main board to
solenoid valve (Air
Intake)
TP24 GND
+24VDC from the
main board to
solenoid valve (Air
Exhaust)
Pin 6 black
+12VDC out to
sensor board from
main board
Pin 2 black
+24VDC out to
compressor from the
main board (variable
voltage) - D65 =
24.4% - D66 = 48.8%
- D67 = 73.2% - D68
= 97.6% - D71 =
110%