No back right deflate
No back left inflate
No back left deflate
No seat right inflate
www.stryker.com
Troubleshooting
Loose or disconnection of the pod
manifold connector
No air is exiting the turn bladder
Loose or disconnection of the pod
manifold connector
Possible leak in the turn bladder
No air entering the turn bladder
Loose or disconnection of the pod
manifold connector
No air is exiting the turn bladder
Loose or disconnection of the pod
manifold connector
Possible leak in the turn bladder
No air entering the turn bladder
2971-109-0 01 REV B
Make sure that the pod manifold
connector is set into the manifold and
secured
Make sure that the LED (D23) is lit to
show the BRD solenoid is energizing
(see
Main power board assembly on
page
1-28)
If energized, replace the BRD inflate
solenoid
Make sure that the pod manifold
connector is set into the manifold and
secured
Listen for air flowing into the turn
bladder
Look for a leak in the turn bladder
If you find a leak, replace the turn
bladder assembly
Make sure that the LED (D34) is lit to
show the BLI solenoid is energizing
(see
Main power board assembly on
page
1-28)
If energized, replace the BLI inflate
solenoid
Make sure that the pod manifold
connector is set into the manifold and
secured
Make sure that the LED (D36) is lit to
show the BLD solenoid is energizing
(see
Main power board assembly on
page
1-28)
If energized, replace the BLD inflate
solenoid
Make sure that the pod manifold
connector is set into the manifold and
secured
Listen for air flowing into the turn
bladder
Look for a leak in the turn bladder
If you find a leak, replace the turn
bladder assembly
Make sure that the LED (D25) is lit to
show the SRI solenoid is energizing
(see
Main power board assembly on
page
1-28)
If energized, replace the SRI inflate
solenoid
E E N N
1-69