Troubleshooting
Loose or disconnection of the right
turn bladder hose quick connection
coupler
Loose or disconnection of the pod
manifold connector
Possible leak in the turn bladder
No air entering the turn bladder
Loose or disconnection of the right
turn bladder hose quick connection
coupler
No air is exiting the turn bladder
Loose or disconnection of the pod
manifold connector
Possible leak in the turn bladder
No air entering the turn bladder
2971-109-001 REV B
Make sure that the right turn bladder
hose quick connection coupler is
together and locked
Make sure that the pod manifold
connector is set into the manifold and
set
Listen for air flowing into the turn
bladder
Look for a leak in the turn bladder
If you find a leak, replace the turn
bladder assembly
Make sure that the LED (D26) is lit to
show the TLI solenoid is energizing
(see
Main power board assembly on
page
1-28)
If energized, replace the TLI inflate
solenoid
Make sure that the right turn bladder
hose quick connection coupler is
together and locked
Make sure that the LED (D27) is lit to
show the TLD solenoid is energizing
(see
Main power board assembly on
page
1-28)
If energized, replace the TLD inflate
solenoid
Make sure that the pod manifold
connector is set into the manifold and
secured
Listen for air flowing into the turn
bladder
Look for a leak in the turn bladder
If you find a leak, replace the turn
bladder assembly
Make sure that the LED (D22) is lit to
show the BRI solenoid is energizing
(see
Main power board assembly on
page
1-28)
If energized, replace the BRI inflate
solenoid