Power supply replacement

E E N N
Tools required:
Zipper pull tool or equivalent
T10 Torx driver
ESD system
Procedure:
1.
See
Foot box cover access on page
2.
Properly ground yourself (see
3.
Remove the wires for the fans and DC power from the wire clip (A).
4.
Disconnect the DC power connector from the main board (AR) (connector R)
5.
Disconnect the AC power connector (B) from the power supply
6.
Disconnect the two foot box fans (D).
7.
Disconnect the ground wire connector from the power supply.
8.
Using a T10 Torx driver, remove the four screws (C) that secure the power supply (AY) to the foot box
on page
1-44). Save the screws.
Note: During reinstall, torque (C) to 11 ± 2 in-lb.
1-44
Service
1-40.
Protecting against Electrostatic Discharge (ESD) on page
Figure 1-20: Power supply
2971-109-001 REV B
1-33).
(Figure 1-20 on page
D
B
A
(Figure 1-20 on page
1-44).
1-44).
(Figure 1-20