Cover Replacement, Top - Stryker Isolibrium 2971 Manuel D'utilisation Et D'entretien

Surface de soutien intégré avec intouch fl27 2131 / 2141 version 4.0
Masquer les pouces Voir aussi pour Isolibrium 2971:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Cover replacement, top

E E N N
Tools required:
Zipper pull tool or equivalent
Procedure:
1.
Apply the InTouch brakes.
2.
Unplug the support surface power cord.
3.
Remove the headboard and footboards.
4.
Lower all the siderails.
5.
Unsnap the two secure snaps (C) at the foot end corners
6.
Unsnap the four corner retainers (B)
7.
Insert a zipper pull tool or equivalent through the hole in the zipper slider to unlock.
8.
Unzip the support surface top cover from the bottom using the zipper pull tool or equivalent (A)
1-34).
Note: The zipper starts on the patient right side near the foot end.
9.
Remove and discard the top cover.
10. Reverse to install the new cover.
Notes
After installation, remove the zipper pull tool or equivalent from the zipper.
Connect the corner retainers by placing the strap below each D-ring and snaps to the cover. Repeat for the
other three corners.
Make sure that the watershed covers the zipper.
11. Verify proper operation before returning the product to service.
1-34
Service
(Figure 1-9 on page
(Figure 1-9 on page
1-34).
A
Figure 1-9: Top cover
2971-109-001 REV B
1-34).
(Figure 1-9 on page
B
C
www.stryker.com