Pump assembly replacement (Continued)
E E N N
Note: During reinstall, torque (A) to 22 ± 4 in-lb.
6.
Move the pump assembly to the side so the area in front of the manifold is clear to remove the hose from the
manifold.
7.
With a firm grasp on the manifold hose, pull straight outward while twisting.
Note: Be careful when removing and installing to not damage to the manifold barb.
8.
Grasp the resonator hose and hold the resonator while pulling outward on the hose.
9.
Remove and discard the pump assembly.
10. Reverse steps to reinstall.
Notes
During reinstall, make sure that you hold the pump springs while tightening so the springs do not turn.
During reinstall, make sure that you check the pump hose is not kinked or bent.
During reinstall, make sure to align the Velcro® patches.
After installation, remove the zipper pull tool or equivalent from the zipper.
Make sure that the watershed covers the zipper.
Do not dispose of as unsorted municipal waste. See your local distributor for return or collection systems
available in your country.
11. Verify proper operation before returning the product to service.