Pump assembly replacement

Tools required:
Zipper pull tool or equivalent
Diagonal pliers
T15 Torx driver
Procedure:
1.
See
Foot box cover access on page
2.
Disconnect the pump power connector from the main board (connector F).
3.
Disconnect the cable from the three wire retainers.
4.
Using diagonal pliers, carefully cut the cable ties where the inlet hose connects to the manifold (BG) and the exhaust
hose connects to the resonator
Note: During installation, before cutting the end off of the cable tie, use the diagonal pliers to pry the slack out of the
cable tie (one to two clicks).
5.
Using a T15 Torx driver, remove the four screws (A) that secure the pump assembly (AJ) to the foot box
24 on page
1-51). Save the screws.
www.stryker.com
Service
1-40.
(Figure 1-24 on page
1-51). Do not cut the manifold barb.