Connect the Blue Fluent In-FloPak
WARNING! Original accessories
For your own safety and that of your patient, use only Fluent accessories.
4a.
Snap the Fluent In-FloPak into place and wait for the green check mark before proceeding.
Ensure Fluent In-FloPak is flushed with front of the console.
4b.
Hang the Fluid Bag.
4c.
Close both tubing clamps.
4d.
Spike the Fluid Bag.
4e.
Connect the hysteroscope Inflow Tube (that has a blue band adjacent to the Luer lock connector)
to the hysteroscope Inflow Channel.
4f.
Connect the light cord and the camera to the hysteroscope.
Note:
Do not unclamp the fluid bag tubes until prompted by system.
Connect the Yellow Fluent Out-FloPak
WARNING! Original Accessories
For your own safety and that of your patient, use only Fluent accessories.
5.
Snap the Fluent Out-FloPak into place and wait for the green check mark before proceeding. Ensure Fluent
Out-FloPak is flushed with front of the console.
6a.
Connect the hysteroscope Outflow Tube (that has a yellow band adjacent to the Luer lock connector) to the
hysteroscope Outflow Channel.
6b.
Firmly press the Waste Tube Connector onto the Waste Bag opening.
6c.
Connect the yellow suction connector onto the Under Buttocks (UB) Drape Port.
7.
Touch 'Next' on the Touchscreen.
After you touch 'Next', the Fluent In-FloPak and Fluent Out-FloPak will lock into place. Unclamp Fluid Bag tube
before priming. If you do this step before Step 7, fluid will leak out of the hysteroscope.
20
Fluent Fluid Management System
Chapter 3: Connecting the System Components
Hysteroscope Inflow Tube
Close the tube clamp and spike.
Hysteroscope Outflow Tube
Waste Tube Connector
English