Hologic FLUENT Manuel D'utilisation page 39

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 329
• Orient the MyoSure Tissue Removal Device cutting window against the tissue when the foot pedal is activated. If you see flashing from the
scope, it means the window is not directly over the tissue.
• Ensure the outflow tubing is not occluded or pinched.
• Avoid bending the MyoSure tissue removal device handle.
Fluent Fluid Management System
• If the Fluent Fluid Management System powers down unexpectedly, leave it off for 15 seconds, restart the system and follow on-screen
prompts.
Display Error and Troubleshooting
Warning Text
Supply bag low -
Waste bag almost full -
Approaching Deficit Limit -
Warning - High Pressure -
Reprime required -
MyoSure device not found -
MyoSure motor stalled -
in-FloPak locking error -
Pump is moving when
calibration pressure is
requested -
English
Fluent Fluid Management System
Warning Code
Description
01010033
Bag contains less than 500ml
01010034
WB will be full in 500ml
01010037
Deficit is within 75% of set limit
010A0004
Any momentary high pressure
condition that lasts more that 2s but
less than 5s
010A0008
Unexpected difference between
measured and predicted uterine
pressure
01070001
Device is not fully plugged in or
defective
01070009
Excessive force or bend is being
applied to the device
01040004
In-FloPak failed to properly lock into
the console, or is not fully seated
01040008
Pump is on during pressure calibration
check
Chapter 10: Troubleshooting
Resolution
Add and spike another Fluid bag
Replace waste bag
Readjust Deficit Limit if appropriate.
Ensure that line between in-FloPak
and scope inflow is open. Ensure
scope tip is not up against uterine
fundus.
During priming, the scope must be
open to the air at the level of the
patient.
Ensure that the fluid flow at the end of
the scope is not blocked.
Scope must not be in the uterus when
priming.
Ensure that inflow tube clamp and
scope stopcock are open to allow fluid
to flow freely.
Once checked per above steps -
Reprime the scope.
Ensure device is properly plugged
in. If problem persists use another
MyoSure device.
Reduce bending force on MyoSure
device, then release and depress foot
pedal to resume cutting. If problem
persists, use another MyoSure device.
Make sure the In-FloPak is fully
seated into the console. You may need
to hold it down during the locking
sequence.
Exit calibration check and run the
calibration check again.
37

Publicité

Table des Matières
loading

Table des Matières