This chapter provides information and instructions for replacing Fluent Fluid Management System Fluid Bags, Waste Bags, and Tissue Traps.
Depending on the procedure, you may need to replace disposable components. To alert you of changes, the Fluent Fluid Management System IV
Pole and Waste Bag built-in scales prompt you if you need to take action. Changing waste bag without pausing the system may lead to inaccurate
deficit values.
Note:
Prior to replacing any System Component, ensure any error messages are resolved first. Failure to do so may result in inaccurate deficit
readings or final procedure results.
If Fluid or Waste Bag are not changed in accordance with the instructions below, the deficit and/or final procedure results may need to be
calculated manually.
Replacing the Fluid Bag
Add a new Fluid Bag when the system alerts you the supply bag is low.
Perform the following steps to replace the Fluid Bag during a procedure:
1.
Hang bag on either hook (do not remove empty bag, doing so may result in inaccurate final procedure results, such as total fluid
volume collected.)
2.
1-3 Liter bag may be used.
3.
Maximum weight on hook = 6 Liters of fluid.
4.
Insert spike into bag port.
5.
If using two bags, only spike one at a time.
6.
Open clamp on the spiked bag before resuming procedure.
Replacing the Waste Bag
Replace the Waste Bag when the system alerts you.
Perform the following steps to replace the Waste Bag during a procedure:
1.
Touch 'Pause' to pause the system. (Not touching 'Pause' may result in inaccurate deficit readings.)
2.
Remove outflow tubing from Waste Bag.
3.
Screw cap into Waste Bag Opening.
4.
Remove from hooks and discard according to facility protocols.
5.
Hang bag by placing both outer rings on both hooks.
6.
Bag should sit clearly in hook notches.
7.
Make sure Waste Bag cap is not closed.
8.
Attach outflow tubing from yellow out-FloPak onto Waste Bag opening.
9.
If 'Missing Waste Bag' error is displayed, make sure Waste Bag is in valley of the hooks, touch 'Clear' and continue with the procedure.
Replacing the Tissue Trap
Perform the following steps to replace the Tissue Trap during a procedure.
1.
Touch 'Pause' to pause the system.
2.
Open Tissue Trap Holder lid.
3.
Remove Tissue Trap and place into pathology container.
4.
Place new Tissue Trap into the Tissue Trap Holder.
5.
Close the Tissue Trap Holder lid securely.
6.
Make sure Waste Bag cap is still attached to Waste Bag.