Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 329
Caution!
Check to ensure the available wall outlet voltage matches the data listed on the label attached to the rear of the system. Incorrect voltage can
cause errors and malfunctions and may destroy the system.
The power connection must be equipped with a grounding outlet. Use the Fluent Fluid Management System Power Cord to establish a connection
between the wall outlet and the Power Port located on the rear of the Fluid Management System.
Note:
For more information about power cord safety, see Chapter 12: Technical Specifications.
Identifying Fluent Flo-Pak Tubes
Fluent In-FloPak Tubes
The tubes of the blue Fluent In-FloPak connect to the following:
• The Fluid bag
• The Hysteroscope Inflow Channel
Fluent Out-FloPak Tubes
The tubes of the yellow Fluent Out-FloPak connect to the following:
• The Waste Bag
• The Under Buttocks (UB) Drape
• The Hysteroscope Outflow Channel
• Optional: The MyoSure Tissue Removal Device (TRD). Use only if performing MyoSure procedure.

System Setup

Perform the System Setup steps in the order that is most logical for your facility. Handle all materials according
to your facility's protocol. Perform the following steps to prepare the Fluent Fluid Management System for use.
Hang the Fluid Bag
WARNING!
The maximum amount of weight should not exceed 6L of fluid. Doing so may negatively impact fluid deficit
accuracy.
1.
Hang up to 6L of fluid from the IV Pole with distension media appropriate for the procedure. If you need to
replace a Fluid Bag during the procedure, see the Chapter 6: Replacing Components.
Note:
The Fluent Fluid Management System does not need to be powered on to hang the Fluid Bags. If it is on, then
follow the prompts. After the Fluid Bags have been hung properly, the system displays a green check mark.
Unpack the Fluent Procedure Kit
2a.
Open the Fluent Procedure Kit package.
2b.
Set the non-sterile Waste Bag aside.
2c.
Peel back the sterile seal to open the Fluent In-FloPak and Fluent Out-FloPak and place them on a
sterile surface.
Hang the Waste Bag
WARNING!
The maximum amount of weight should not exceed one full Waste Bag (6kg). This can negatively impact fluid deficit accuracy.
3.
Hang a new Waste Bag on the Waste Bag Hanger to begin any new procedure.
After the Waste Bag has been hung properly, the system displays a green check mark.
English
Fluent Fluid Management System
Chapter 3: Connecting the System Components
Waste Bag Hanger
19

Publicité

Table des Matières
loading

Table des Matières