9.4 CLEANING-DISINFECTION-STERILIZATION-STORAGE OF THE FLEXIBLE HOSE

The flexible hose must be disconnected from the pneumatic turbine and from the generator, cleaned,
decontaminated, sterilized, and dried after each intervention.
Steps What to do
1
Cleaning
2
Disinfecting
3
Rinse and dry
4
Sterilization
5
Storage
12
MORIA
#65005-F-02.2017
IMPORTANT WARNINGS
➢ Do not immerse.
➢ The flexible hose can be cleaned on the outside with a lint-free
cloth soaked with alcohol.
➢ Do not immerse.
➢ The flexible hose can be decontaminated on the outside with a
decontaminating solution.
➢ The flexible hose must be rinsed on the outside with a lint-free
disposable cleaner soaked in sterile water.
➢ It must be then dried with a disposable lint-free cleaner.
➢ The flexible hose can be sterilized in the autoclave 134°C for 18
minutes.
➢ Sterilization in USA only: sterilize with an autoclave (steam
sterilizer with a vacuum phase) at a minimum of 132°C for at
least 4 minutes.
➢ It must be removed from the autoclave immediately after
sterilization.
➢ Never use the flexible hose immediately after sterilization; let it
cool at room temperature.
➢ The flexible hose must be kept in a dry place, away from dust.
➢ In order to avoid that the hose get damaged, it is important to
store it carefully in its box (Figure L).