Les langues disponibles

  • FRANÇAIS, page 18
5
Assemble
pneumatic
turbine hose to
CB turbine and
console unit
6
Check for blade
oscillation
7
Select suction
ring according to
nomogram, then
inspect it
8
Connect the
suction ring to
the aspiration
tubing

7 OPERATION

Steps
What to do
1
Preliminary check before any usage
2
Connect the CB turbine to the control
unit. Check the pressure delivered by
the console at the turbine (operating
range between 2.5 and 3.5 bars).
3
Assemble the head in the suction ring
4
Check for blade oscillation
5
Position the suction ring on the eye.
Activate vacuum by pressing the
"Vacuum" footswitch once.
• Check that the gaskets on the turbine connector and on the control unit
connector are present and not damaged.
• Check the turbine hose for cracks and possible damage. Replace if
necessary.
• Connect the pneumatic turbine hose to the back of the turbine.
• Screw the metal collar of the turbine hose onto the back of the turbine
motor. The connector shape allows only one possibility of connection and
prevents from wrong assembly. Finger tighten only.
• Place the turbine connector hose onto the control unit front panel
connector.
• The connector shape allows only one possibility of connection and
prevents from wrong assembly.
• Refer to EVOLUTION 2, 3 or 3E user manual for operations prior use.
• Activate CB turbine to check for smooth, strong and uninterrupted blade
oscillation.
• Do not use the CB turbine if the blade oscillation is not smooth, regular and
uninterrupted.
• Refer to the indicative nomogram for the selection of the ring size.
• The suction ring must be carefully inspected prior to use; it must be
perfectly clean, free from debris, particles and deposits.
• Carefully check that the suction port is not obstructed.
• The hinge will be placed facing the arrow on the suction ring.
• The aspiration tubing is supplied sterile and for single use only.
• Check that package is undamaged, unopened and within the expiration
date.
• Inspect the aspiration tubing and in case of kinks or obstructions, replace it.
• It is imperative to verify that the devices are perfectly
clean and free of any organic or other residues.
• See EVOLUTION 2, 3 or 3E user manual for
operations prior use.
• The CB-CBm-CBSU head pivot opening should fit
easily over the pivot post of the CB suction ring.
• Align suction ring pivot post and CB head pivot hole.
Carefully place the assembled CB keratome and CB
turbine motor onto the suction ring.
• Avoid inadvertent contact with suction ring pivot post
and CB head blade area. If contact occurs, blade edge
may become damaged, and a new blade should be
used.
• Refer to EVOLUTION 2, 3 or 3E user manual for
operations prior use.
• Activate CB turbine to check for smooth, strong and
uninterrupted blade oscillation.
• Do not use the CB turbine if the blade oscillation is not
smooth, regular and uninterrupted.
• Verify that the device will not interfere with any other
medical devices used during the procedure.
IMPORTANT WARNINGS
#65005-F-02.2017
MORIA
H
I
J
Related
picture
H
K