Troubleshooting; Care & Maintenance; Introduction; Initial Treatment And Storage - MORIA 65005 Manuel Utilisateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
6
Check the intraocular pressure with
the tonometer
7
Lubricate the suction ring, the
head and the blade
8
Once the system is in position,
activate blade oscillation by pressing
the "Forward" footswitch.
Manually rotate the CB
microkeratome within the suction ring
until the stop of the suction
ring is engaged.
Stop the turbine motor and manually
reverse the CB microkeratome.
9
Release the "Vacuum" footswitch by
pressing once.
Remove the suction ring.

8 TROUBLESHOOTING

For any further information and description, please refer to EVOLUTION 2, 3 or 3E console user manual.
9 CARE & MAINTENANCE
The recommendations for cleaning, disinfection and sterilization are updated and available on MORIA's
restricted-access website: http://www.moria-surgical.com. You must register to access it.

9.1 INTRODUCTION

In case of unusual vibrations or noises, do not use the unit and contact your distributor.
Please contact MORIA for any other servicing.
The following are only recommendations. They must be adapted according to the laws in the country
of utilization. All the products and solutions for cleaning and disinfection must be used according to the
manufacturer's instructions.
MORIA recommends
do not touch potentially contaminated areas,
use gloves during cleaning and decontamination operations.

9.2 INITIAL TREATMENT AND STORAGE

To avoid risks of condensation inside the packaging, unpack and store the devices in a clean, dry environment.
Do not store the instruments in an environment or next to other products that may possibly have a corrosive
or magnetic effect. Avoid any contact between devices, especially those comprised of different materials.
The device should be damage free and have no scratches or other surface defects.
Fragile devices or those requiring particular manipulation must be handled separately, taking special care to
protect delicate parts.
10
MORIA
#65005-F-02.2017
• The tonometer must be perfectly dry and used only on
dry eyes.
• If the pressure is below 65 mm Hg, do not proceed
with the surgery.
• Prior to use, lubricate the rails of suction ring,
head, and blade with balanced salt solution or
another appropriate ophthalmic solution.
• The device should not be placed in contact with
solutions for which the potential interactions are
unknown.
• See EVOLUTION 2, 3 or 3E console user manual.
• Hold the suction ring by its handle; check that your
fingers do not hinder the forward movement of the
device.
• Check nothing prevents or modifies head movement in
the rails of the suction ring
• Ensure that there are no obstructions (speculum,
eyelashes, eyelids, etc.) in its path.
• Discard the disposable blade (CBm) or disposable
head (CBSU) and aspiration tubing after each
procedure in an appropriate container.