9.3 CLEANING-DISINFECTION-STERILIZATION-STORAGE OF THE TURBINE

The pneumatic turbine must be disconnected from its hose.
It must be cleaned, disinfected, dried and sterilized after each surgical procedure.
ACTIONS
1. Manual external cleaning
• After use, clean the outside of the turbine with a lint-free
cloth dampened with a cleaning solution (eg: Alkazyme
/ Laboratoires Alkapharm) to remove any residue.
2. Manual external disinfection
• After cleaning, dry every part of the turbine with a lint-
free cloth dampened with a disinfecting solution.
3. Lubrication then cleaning before sterilization
• Lubricate the handpiece using a KaVo spray only
and the appropriate nozzle before every autoclave
sterilization cycle. A single jet of KaVo spray is sufficient
to lubricate the turbine while avoiding excess lubricant
inside the turbine.
• Clean the turbine with a lint-free cloth then keep it
upside down for 20 minutes before the autoclave
sterilization to expel excess lubricant.
4. Sterilization
• Place the turbine and its hose in the dedicated
sterilization box.
• Sterilize the turbine in the autoclave (134°C – 273°F) for
18 minutes.
Sterilization in USA only: sterilize with an autoclave
(steam sterilizer with a vacuum phase) at a minimum of
132°C for at least 4 minutes.
5. Storage
• Once this sterilization cycle is completed, remove the
handpiece as promptly as possible and store it upright
until it cools down and is dry.
• Once cooled down and dry, the handpiece can be
stored in an enclosed dry place.
IMPORTANT WARNINGS
• Do not use abrasives or scrapers to clean items. It
could damage their usability.
• Do not immerse the turbine in any solution whatsoever,
®
including water.
• Cleaning is not disinfecting! Disinfection is a separate
step described in point 2.
• Use appropriate disinfection solutions in accordance
with the manufacturer's instructions.
• Do not immerse the turbine in any solution whatsoever,
including water.
• Use the recommended spray. Do not use other lubricant
sprays, vaseline or grease.
• Shake the spray before using.
• Place the spray nozzle on the smallest hole of the
turbine and press once only.
• KaVo spray (ref 411-9630) is not sold by MORIA. It can
be purchased from the Kavo website (www.kavo.com).
• Remove the turbine from the sterilizer immediately after
the sterilization procedure.
• Never use the turbine immediately after sterilization. Let
it cool down to room temperature first.
• To avoid oxidation, make sure that the turbine is
perfectly dry before putting it in the storage box.
• While it is not being used, the CB turbine must be kept
in its storage box in a dry atmosphere (Photo L).
• Before use, check that the handpiece has no lubricant
on it. If it has, do not use it. Do not lubricate before use.