Percutaneous Leads Directions for Use
11. For Permanent IPG Implantation, if extensions are used, proceed to the instructions for "Connecting
the Lead Extension" on page 16.
12. For Permanent IPG Implantation, if splitters are used, proceed to the instructions for "Lead Connection
to Splitter" on page 9.
13. For permanent IPG implantation, refer to the instructions as follows:
SCS System
Precision SCS System
Precision SCS System with
MultiWave Technology
Precision Spectra SCS System
Precision Novi SCS System
Note: The following Codman Disposable Catheter Passers may be used in place of the Boston
Scientific tunneling tool:
REF 82-1515 (36 cm); REF 82-1516 (55 cm); REF 82-1517 (65 cm)
Note: When using a Codman Disposable Catheter Passer, tunnel from the IPG pocket to the
midline incision using the standard technique.
Connecting the Lead Extension
1. Wipe clean the proximal end of the lead, then insert the proximal end into the lead extension connector
or splitter until it stops and the retention ring (long ring) is under the setscrew.
Note: If there appears to be an obstruction when inserting the lead into the lead extension
connector, use the hex wrench to loosen (counterclockwise) the setscrew and/or gently
rotate the lead to help advance the proximal end.
2. Ensure that the lead is fully inserted before tightening the set screw to prevent lead damage.
3. Using the hex wrench supplied, turn the extension connector setscrew clockwise until it clicks,
indicating lock.
Note: Ensure that the hex wrench is fully seated in the setscrew before tightening.
Note: The hex wrench is torque-limited and cannot be overtightened.
4. Form an appropriately-sized pocket using blunt dissection on either side of midline for coiled excess
lead and extension connectors.
Percutaneous Leads Directions for Use
90970887-03 Rev A
16 of 441
Reference Document for instructions
Precision Spinal Cord Stimulator Clinician Manual
Precision Spinal Cord Stimulator System with MultiWave
Technology Clinician Manual
Precision Spectra IPG DFU
Precision Novi IPG DFU
Section Title
"IPG Implantation"
"IPG Implantation"
"IPG Implantation"
"IPG Implantation"