Implantable Pulse Generator
7. Unscrew and remove the tunneling tool handle.
8. Grasp the tip of the tool with one hand while holding the straw in place with the other hand. Pull the
tunneling tool shaft out through the straw.
9. Push the lead or extension through the straw, then withdraw the straw.
10. Pull the proximal end(s) out of the exit point.
11. Wipe the proximal end(s) clean.
CAUTION: Do not tunnel splitter.
Note: The following Codman Disposable Catheter Passers may be used in place of the Boston Scientific
tunneling tool for the 2x8 Surgical Lead, Linear™ 8 Contact Leads and Infinion 16 Lead only:
REF 82-1515 (36 cm); REF 82-1516 (55 cm); REF 82-1517 (65 cm)
Note: When using a Codman Disposable Catheter Passer, tunnel from the midline incision to the IPG
pocket using the standard technique.
Implantable Pulse Generator
91053242-08 Rev A
8 of 279
Production
Boston Scientific Confidential. Unauthorized use is prohibited.
Montage MRI IPG Entrada 2.0 19L
92070454 A.1
Page 12 of 284