Remote interface triggering
Immediate triggering
In the immediate trigger mode, the trigger signal is always present. When you
place the multimeter in the wait-for-trigger state, the trigger is issued
immediately. This is the default trigger source for the U3606B.
The immediate trigger mode is available from the remote interface only.
NOTE
Remote interface operation:
– The TRIGger:SOURce IMMediate command selects the immediate trigger
– After selecting the trigger source, you must place the instrument in the
Refer to the U3606B Programmer's Reference for the syntax and complete
descriptions for these commands.
Software (bus) triggering
The bus trigger mode corresponds in function as the
except the trigger is initiated by sending a bus trigger command, after selecting
BUS as the trigger source.
The bus trigger mode is available from the remote interface only.
NOTE
Keysight U3606B User's Guide
source.
"wait-for-trigger" state using the INITiate[:IMMediate] or READ?
command. A trigger will not be accepted from the selected trigger source until
the instrument is in the "wait-for-trigger" state.
Operation and Features
Single triggering
mode,
2