The U3606B is equipped with ramp and scan capability. Use the ramp function to
generate a ramp signal with its end amplitude position and number of steps based
on the preset input parameters and the scan function to generate a scan signal
with its end amplitude position, step dwelling time, and number of steps based on
the preset input parameters.
Ramp signal
A typical ramp signal length is based on the following parameters:
– the amplitude end position, and
– the number of steps required to reach the amplitude end position.
You can configure the ramp signal amplitude end position and number of steps in
the Utility menu.
The ramp dwelling time will be set to the fastest of the instrument capability
(typically ~100 ms per step). A higher number of steps provides a more linear
ramp signal. This however will result in an increase in the total dwelling time.
A lower number of steps will result in a shorter total dwelling time and a more
stepped ramp signal.
Keysight U3606B User's Guide
Amplitude
(V/A)
Amplitude
end position
Number
of steps
0
DC Power Supply Operation
Total dwelling time
3
Time (t)
125