4
System-Related Operation
Configuring the data logging parameters
The data logging function provides you the convenience of recording test data for
future review or analysis. Since data is stored in the non-volatile memory, the data
remains saved even when the U3606B is turned OFF.
The data logging feature collects measurement information over a user-specified
duration.
Press [ ] or [ ] to change the data logging loop number (1 to 29000).
Press [ ] or [ ] until the cursor is position on the data logging interval. Press [ ]
or [ ] to change the data logging interval (1 to 99999 s).
– You can record up to 28800 data (8 hours × 60 minutes × 60 seconds).
NOTE
– The data logging operation will stop automatically when it is completed or
Recording measurement data (data logging)
Press [ ] or [ ] to switch between "StoP", "Cont", or "StArt".
Select "StArt" to start the data logging operation, select "StoP" to stop the data
logging operation, and select "Cont" to resume a data logging operation from
where it was last stopped.
148
when the U3606B's memory is full.