4
System-Related Operation
Setting the output protection state
The OVP and OCP features depend on the output protection state function. The
output protection state must be enabled for the OVP and OCP features to be
active in the circuit. Disabling the output protection state will deactivate both the
OVP and OCP features even if a trip level is set for either the OVP or OCP.
The output protection state does not affect the over-voltage and over-current
NOTE
limits.
– Disabling the output protection state may result in equipment damage if
CAUTION
– In the event that there is a large current over the system protection
Table 4-4
System protection values
Range
S1 (30 V/1 A)
S1m (30 V/100 mA)
S2 (8 V/3 A)
S2m (1000 mV/3 A)
S1S2 (autoranging)
158
an over-voltage or over-current condition occurs. It is recommended to
have the output protection state enabled.
values
(Table
4-4), the U3606B will trip to protect itself even if the triP
option in Utility menu is set to "no".
System protection (approx.)
1.3 A
230 mA
3.5 A
3.5 A
3.5 A
Keysight U3606B User's Guide