Probe the test points and read the display
Current measurement summary
Table 2-2
Current measurement summary
Item
DC current measurement
Available ranges
Measurement method
Burden voltage and shunt
resistance
Input protection
Keysight U3606B User's Guide
In this example, we obtain an AC current measurement of 1.2345 A,
with autoranging enabled and 5½ digit resolution selected by default.
Description
10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A
Sigma Delta A-to-D converter
– < 0.2 V, 10 Ω for 10 mA range
– < 0.2 V, 1 Ω for 100 mA range
– < 0.3 V, 0.05 Ω for 1 A range
– < 0.7 V, 0.05 Ω for 3 A range
Protected with a 3.15 A/500 V, FF fuse