Remote Operation
The U3606B is shipped with both a GPIB (IEEE-488) interface and a USB 2.0
interface on the rear panel. Only one interface can be enabled at a time. The GPIB
interface is selected by default when the U3606B is shipped from the factory.
The remote interface can be selected from the front-panel only.
– The interface selection is stored in the non-volatile memory, and does not
– If you select the GPIB interface, you must select a unique address for the
– You can select between two USB class: USB-TMC or USB-CDC. USB-TMC is
The instrument automatically enters the remote state whenever SCPI commands
are received over the GPIB or USB interface. The Rmt annunciator is illuminated
and the front panel keys are locked when in the remote state. Press [Local] to
return the U3606B to front panel operation.
Keysight U3606B User's Guide
change when power is turned off or after a remote interface reset.
U3606B. The current address for the U3606B is displayed on the lower
secondary display in the Utility menu.
the default USB 2.0 full-speed communication protocol interface that is
compliant to USB standards, while USB-CDC is simulates a serial
communication (RS-232) interface at the PC through the physical USB port
connection.