3
DC Power Supply Operation
– After adjusting the OC value, you can also press [Limit] again to save the
NOTE
– The OC feature is always enabled by default whenever you select constant
– If the OC value is set to a greater value than the OCP value, the OCP value
– If the OC value is set to zero, this will result in the output voltage being
– The OC value is not applicable for square-wave output.
– The OC value is limited by the range selected. Press [Shift] > [Range] to
Table 3-4
CV mode over-current range and values
Range
S1 (30 V/1 A)
S2 (8 V/3 A)
S2m (1000 mV/3 A)
S1S2 (autoranging)
[a] If the S1S2 (autoranging) range is selected, you will not be able to adjust the OC value.
118
changes made or press [Shift] > [Exit] to discard the changes made.
current mode. You cannot disable the OC feature.
will be adjusted to equal the OC value.
dropped to zero for limiting.
select an appropriate range. You can only select a range when the output is
disabled (the SBY annunciator is illuminated).
Over-current value
Minimum
0 A
0 A
0 A
N/A
Maximum
1.05 A
3.15 A
3.15 A
[a]
Keysight U3606B User's Guide