2
Operation and Features
Table 2-2
Current measurement summary (continued)
Item
AC current measurement
Available ranges
Measurement method
Burden voltage and shunt
resistance
Input protection
Crest factor
Peak input
Please expect some delay when obtaining a stable AC current reading on the
NOTE
following test conditions:
– The signal has a large DC current offset comparable to the AC current signal
– Changes on the range/function
Using the AC+DC function
NOTE
The U3606B is a true-rms multimeter that returns an accurate rms reading not
only for sine waves, but also other AC signals such as square, triangle, and
staircase waveforms without any DC offset. However, you may choose to return
the measured AC signal with its DC offset by using the AC+DC function.
52
Description
10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A
AC coupled true rms
– < 0.2 V/10 Ω for 10 mA range
– < 0.2 V/1 Ω for 100 mA range
– < 0.3 V/0.05 Ω for 1 A range
– < 0.7 V/0.05 Ω for 3 A range
Protected with a 3.15 A/500 V, FF fuse
For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at
full scale.
300% of range. Limited by maximum input.