Disconnect circuit power and discharge all high-voltage capacitors before
CAUTION
measuring resistance or conductance, or testing circuit continuity, to avoid
damaging the U3606B or the device under test.
DC power supply functions are locked
NOTE
The DC power supply functions are locked when you select the low-resistance
(Lo Ω) measurements. You cannot use the DC power supply module
simultaneously with the digital multimeter module when you are measuring
4-wire low-resistance measurements.
To unlock the DC power supply functions, exit the low-resistance (Lo Ω)
measurement by selecting another multimeter measurement.
Delayed response for low-resistance readings
NOTE
For measuring low-resistance, a delayed response should be expected from the
front panel display. For remote interface operation, increase the SCPI query
time-out value (typically 15000 ms).
Low-resistance measurement summary
Table 2-5
Low-resistance measurement summary
Item
Available ranges
Measurement method
Input protection
Keysight U3606B User's Guide
Description
100 mΩ, 1000 mΩ, 10 Ω, 100 Ω, 1000 Ω
Four-wire, the test current is sent from the FORCE terminals and the resistance is measured by
the SENSE terminals.
– FORCE terminals: Protected with a 3.15 A/250 V FF fuse
– SENSE terminals: 1000 V
on all ranges, < 0.3 A short circuit
rms