System-Related Operation
To set the fluctuation count, press [ ] until the cursor position is on the
fluctuation count at the secondary display. Press [ ] or [ ] to change the cursor
position and [ ] or [ ] to change the value for the fluctuation count (0.0% to
9.9%).
– The range of values for the fluctuation count is from 0.0% to 9.9%, with the
NOTE
– If the input raw signal readings are higher or lower than the fluctuation count
– The smooth function will also reset if the range of the signal changes when
– You can also manually restart the smooth function by pressing [MinMax] for
– Setting the fluctuation count to 0.0% will disable the fluctuation count. It is
To set the number of points, press [ ] until the cursor position is on the number
of points on the primary display. Press [ ] or [ ] to change the cursor position
and [ ] or [ ] to change the values for the number of points (2 to 1999).
152
default value at 1.0%.
of the previous smoothed signal readings, the smooth function calculations
will reset.
For example, a smooth function with a 5 V signal and a fluctuation count of
1.0% will only smooth the input raw signal to the existing smooth calculation
if the input signal is between 4.9 V and 5.1 V. The next input signal will then
be compared to the newly smoothed signal. If the input signal is out of the set
fluctuation count range of the smoothed signal, the smooth function
calculation will reset and restart at the out-of-range signal.
the smooth calculation is running. For example, when the signal range jumps
to the 100 V range from the 10 V range.
more than one second.
recommended that the fluctuation count is enabled.
Keysight U3606B User's Guide