Press [ ] or [ ] to change the scan signal amplitude end position.
Press [ ] or [ ] until the cursor is position on the scan signal number of steps.
Press [ ] or [ ] to change the number of steps for the scan signal to increment
from zero to the amplitude end position.
Press [ ] or [ ] until the cursor is position on the scan signal dwelling time.
Press [ ] or [ ] to change the scan signal dwelling time.
– The increment of each step in the scan signal will be the amplitude end
NOTE
– The scan signal will "dwell" in the present step for the length of time stated in
Keysight U3606B User's Guide
CV/CC scan signal parameters
Amplitude end position
(NN.NNN V or N.NNNN A)
position divided by the number of steps. For example, a 15 V amplitude end
position divided by 100 steps gives an increment of 0.15 V per step.
the scan dwelling time before incrementing to the next step.
System-Related Operation
Scan signal range
Number of
Scan dwelling
steps (NNN)
time (NN s)
4
155