DC Power Supply Operation
Error number 511, "Current output over protection" will be recorded in the error
NOTE
queue when the OCP circuit trips. Enter the Utility menu to read and clear the
error message.
Resetting the OCP condition
Use one of the following methods to reset the OCP circuit after it activates. If the
condition that caused the over-current shutdown is still present, the OCP circuit
will turn the output off again.
– When the OCP circuit trips, the U3606B will immediately prompt you to
– Alternatively, press [Shift] > [Exit] to exit the edit mode without changing the
– You may read and clear the error message by accessing the Utility menu.
If the OCP continues to trip, try lowering the regulated output current, or raising
the OCP trip level.
To protect itself, the U3606B will trip when
NOTE
protection values
the Utility menu.
Over-voltage limit (OV)
The over-voltage limit prevents the output voltage present across the load to
change beyond the programmed over-voltage limit. If the load effect exceeds the
programmed over-voltage limit, the CC output will be lowered to maintain the
output power across the load. The combination of the OV and OVP features create
a closed loop circuit protection for sensitive load behaviors.
112
change the OCP trip level. Use the arrow keys to select a higher OCP trip level
and press [Shift] > [Save] or [Protect] to save the changes.
OCP trip level.
) is detected, even is the triP option
a large current (over the system
(page
158) is disabled in
Keysight U3606B User's Guide