Guidelines and Manufacturer's Statement - Electromagnetic In-
terference Immunity
The Aquilex Fluid Control System is intended for use in an electromagnetic envi-
ronment as described below. The user/operator of the Aquilex Fluid Control Sys-
tem should make sure the device is operated within such an environment.
Electromagnetic
Test level
interference
immunity tests
Discharge of static
± 6 kV contact
electricity (ESD)
discharge
according to
± 8 kV air dis-
IEC 61000-4-2
charge
Electrical fast
± 2 kV for AC
transients / bursts
power lines
according to
± 1 kV for input
IEC 61000-4-4
and output lines
Surges according
± 1 kV normal
to IEC 61000-4-5
mode voltage,
± 2 kV common
mode voltage
Blackouts, brown-
< 5% U
* (> 95%
T
outs, and fluctua-
dip in the U
tions of the power
½ cycle
supply according
40% U
(60% dip
to IEC 61000-4-11
T
in the U
) for 5
T
cycles.
70% U
(30% dip
T
in the U
) for 25
T
cycles.
< 5% U
(> 95%
T
dip in the U
5 s
Supply frequency
3 A/m
magnetic field
(50/60 Hz) accord-
ing to IEC 61000-
4-8
*Note: U
is the mains alternating voltage before applying the test levels.
T
Guidelines and manufacturer's statement - electromagnetic compatibility
Compliance
Electromagnetic envi-
ronment guidelines
In compli-
Floors should be made
ance
from wood or concrete
or covered with ceramic
tiles. If the floor cover-
ing consists of synthetic
material, the relative
humidity should be at
least 30%.
In compli-
The quality of the sup-
ance
ply voltage should be
the same as the voltage
of a typical business or
hospital environment.
In compli-
The quality of the sup-
ance
ply voltage should be
the same as the voltage
of a typical business or
hospital environment.
In compli-
The quality of the sup-
) for
ance
ply voltage should be
T
the same as the voltage
of a typical business or
hospital environment.
If the user/operator of
system requires the
continuation of func-
tionality after power
interruptions/disrup-
tions, it is recom-
mended to supply the
)for
device with power from
T
an uninterruptible
power supply.
In compli-
Magnetic fields of the
ance
mains power fre-
quency should comply
with the typical values
of business and hospi-
tal environments.
EN
39