Guidelines and Manufacturer's Statement - Electromagnetic Compatibility
en
46
12.3
Guidelines and Manufacturer's Statement/Electromagnetic In-
terference Immunity
TheAquilex Fluid Control System is intended for use in an electromagnetic envi-
ronment as described below. The user/operator of the Aquilex Fluid Control Sys-
tem must make sure the device is operated within such an environment.
Electromagnetic
IEC 60601-1-2
interference
test level
immunity tests
Discharge of static
± 8 kV contact
electricity (ESD)
± 15 kV air
according to
IEC 61000-4-2
Electrical fast
± 2 kV for power
transients/bursts
lines
according to
± 1 kV for input
IEC 61000-4-4
and output lines
Modulation
100 KHz
Voltage surges
± 1 kV line(s) to
according to
line(s) ± 2 kV
IEC 61000-4-5
line(s) to earth
Blackouts, brown-
< 5% U
* (> 95%
T
outs, and fluctua-
dip in the U
tions of the power
½ cycle
supply according
< 40% U
to IEC 61000-4-11
T
dip in the U
5 cycles
< 70% U
T
dip in the U
25 cycles
< 5% U
(> 95%
T
dip in the U
5 cycles
Supply frequency
30 A/m
magnetic field
(50/60 Hz) accord-
ing to
IEC 61000 4-8
*Note: U
is the mains alternating voltage before applying the test levels.
T
Compliance
Electromagnetic environ-
Levels
ment/guidelines
± 8 kV con-
Floors should be made
tact ± 15 kV
from wood or concrete or
air
covered with ceramic
tiles. If the floor covering
consists of synthetic
material, the relative
humidity should be at
least 30%.
± 2 kV for
The quality of the supply
power lines
voltage should be the
± 1 kV for
same as the voltage of a
input and
typical business or hospi-
output lines
tal environment.
Modulation
100 KHz
± 1 kV line(s)
The quality of the supply
to line(s)
voltage should be the
± 2 kV line(s)
same as the voltage of a
to earth
typical business or hospi-
tal environment.
< 5% U
*
The quality of the supply
T
) for
(> 95% dip in
voltage should be the
T
the UT) for 1/
same as the voltage of a
2 cycle 40%
typical business or hospi-
(> 60%
U
(60% dip
tal environment. If the
T
) for
in the U
) for
user/operator of the sys-
T
T
5 cycles 70%
tem requires the contin-
U
(30% dip
uation of functionality
(> 30%
T
in the U
) for
after power interrup-
) for
T
T
25 cycles <
tions/disruptions, it is
5% U
(> 95%
recommended to supply
T
dip in the U
)
the device with power
T
) for
for 5 seconds
from an uninterruptible
T
0% U
; 1/2
power supply.
T
cycle at 0°,
45°, 90°,
135°, 180°,
225°,270°,
and 315°0%
U
; 1 cycle
T
and 70% U
;
T
25/30 cycles
Single phase:
at 0°0% U
;
T
250/300
cycles
30 A/m
Magnetic fields of the
mains power frequency
should comply with the
typical values of business
and hospital environ-
ments.