Changing the halogen lamp in the illumination
Disconnect from power supply.
Screw out the Allen screw at the back and
remove the lamp unit from the lamphousing.
Take the lamp out of the socket and replace,
making sure that the contact paths of the lamp
lie on the contacts in the socket.
Do not touch the lamp bulb with your fingers due
to the danger of perspiration burning in.
After the lamp unit has been replaced in the
lamphousing, the lamp holder can be adjusted
vertically by about 2 mm with the Allen screw
from beneath.
Looking through the microscope eyepiece,
adjust the lamp to the height where the greatest
image brightness is achieved.
Macroscopy device
This consists of the reflection optics (35.3), the
macro adapter (35.5) and the macrodual zoom.
Assembly of the macro device
Screw the reflection optics (35.3) onto the tube
flange with the coupling ring (35.2).
Align the macro adapter (35.5) against the
macrodual zoom and secure with the threaded
ring (35.6).
Fasten the macro adapter and the macrodual
zoom to the reflection optics with the coupling
ring (35.4). Watch the guide pin.
Fig. 35 Macro device on the HC FSA 25 PE tube
1 Tube flange, 2 Coupling ring, 3 Reflection optics, 4 Coupling
ring, 5 Macro adapter, 6 Threaded ring, 7 Zoom setting
ring 1 : 4, 8 Zoom factor scale, 9 Scale for magnification factor
of the working distance, 10 Scale for distance of object from
the lower edge of the mirror housing, 11 Mirror housing
1 2 3 4
5 6 7 8 9 10 11