Remove the front cover of the microscope
(22.12); it is no longer required.
Using the supplied 3 mm screwdriver unscrew
the 4 fixing screws (22.1) and remove the cover
with built in tube optics from the microscope.
Fig. 22* Retrofitting the incident light axis (only for BF, DF
and fluorescence! Pol and ICR components can only be
retrofitted by the Technical service)
1 Cover plate (tube optics) with 4 fixing screws, 2 RF 4 in-
cident light module with 4 fixing screws, 3 Lamp mount (with
or without reflector), 4 Mount for switch rod (for switchable
mirror only), 5 Ground glass screen for lamp centration,
6 Analyser mount, 7 3 control points for assembly, 8 Cover
plate or filter box, 9 Adjustment aid (reflector), 10 Lateral
cover plate with 4 fixing screws, 11 Analyser fixture (only
before conversion), 12 Front cover with hole
1
2
4
3
5
3
28
!
Store upside down so as not to damage the
optics. Protect from dust!
Using the crosstip screwdriver, unscrew the 4
fixing screws (22.11) of the analyser mount and
remove it (this component will not be required
again as the analyser mount is integrated in the
HC RF 4 IL module, 22.6).
Using the 2 mm screwdriver, unscrew the 4 Al-
len screws on the lateral cover plate (22.10). This
plate is no longer needed. Please keep the Allen
screws.
6
7
9
8
10
7
11
12
Caution: