Lamphousing 106 z, Hg- and Xe lamps
Attention:
Always insert burner so that
1. the lettering on the metal base is up right
after insertion (different diameters of the
metal base for the Hg 100 and Xe 75 burners
ensure that these are always inserted the
right way up). If one of the bases is labelled
"Up", it must therefore be assembled at the
top .
2. If the lamp bulb has a seal point (7.2), turn
the burner so that this point will be at the
side, not in the light path.
Apart from the halogen lamp the following gas
discharge lamps can be used, all requiring dif-
ferent lamp holders (Fig. 7) and power units:
Type
Hg ultra high pressure lamp 50 W (alternating current)
Xe high pressure lamp
75 W (direct current, stabilized)
Hg ultra high pressure lamp 100 W (direct current, stabilized/non- stabilized) 200 h
Hg ultra high pressure lamp 100 W (direct current, stabilized/
Put the upper pin of the burner between the
clamps of the flexible power supply and clamp
with screw (7.1).
Unscrew the stud (7.3) in the holder slightly,
insert the lower end of the metal base and
retighten the stud.
Exchanging the collector on lamphousing 106 z:
Move the collector to the rearmost position with
the focusing knob (48.19). Pull the focusing knob
of the collector outwards. The collector can
now be removed.
Average life
non-stabilized, type 103 W/2)
!
Make sure that the lamp base and the power
unit have the same number. If the lamp base is
marked L1, for example, L1 must also be set on
the power unit to make full use of the lamp and
not to shorten its life.
Move the collector to the front position with the
focusing knob (48.19).
Remove the protective covering from the
burner (7.7).
Put the lamp holder with burner inserted into the
lamphousing and secure with the screws (8.9).
Try moving the collector (48.19): it must not
touch the power lead.
!
100 h
When closing the lamphousing make sure that
400 h
the pins of the cut-out plug engage in the
sockets (8.8). Retighten the screws of the lid.
300 h
Push the cut-out plug in as far as it will go.
Attach the lamphousing to the microscope
(page 16) and connect to the power unit
(compare mains voltage!).