Pump Pedal Replacement (Hydraulic Base); Uni-Lower Pedal Replacement (Hydraulic Base) - Stryker CUB Pediatric Crib Manuel De Maintenance

Masquer les pouces Voir aussi pour CUB Pediatric Crib:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

PUMP PEDAL REPLACEMENT (HYDRAULIC BASE)

TOOLS REQUIRED:
• 3/16 in. Punch
• Hammer
PROCEDURE:
1.
Raise the litter and the rails to the highest position and apply the brakes.
2.
Using a 3/16 in. punch and hammer, remove the spring pin (A) holding the pump pedal to the pedal rod.
3.
Remove the pump pedal from the pedal rod. If necessary, tap the pump pedal with a hammer to ease removal.
4.
Align the holes on the new pump pedal and the pedal rod and fit the pump pedal on the pedal rod.
5.
Using a hammer, drive the spring pin in until it is flush with the top of the pedal.
6.
Verify proper operation of the pump pedal before returning the bed to service.

UNI-LOWER PEDAL REPLACEMENT (HYDRAULIC BASE)

TOOLS REQUIRED:
• Drill with 3/16 in. Drill Bit
• Floor Jack
• Pop Rivet Tool
PROCEDURE:
1.
Raise the litter and the rails to the highest position and apply
the brakes.
2.
Using a floor jack, raise the side of the stretcher with the
pedal needing repair approximately 4 in. off the floor.
3.
Using a drill with a 3/16 in. drill bit, drill out the two rivets and two washers (B) holding the pedal to the pedal rod.
4.
Align the holes on the new pump pedal and the pedal rod and fit the pump pedal on the pedal rod.
5.
Using a pop rivet tool, install the new uni-lower pedal.
6.
Verify proper operation of the uni-lower pedal before returning the bed to service.
www.stryker.com
Service
A
1900-009-002 REV B
HEAD
Figure 12
B
Return To Table of Contents
FOOT
1-41

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Fl19f/h

Table des Matières