CPR EMERGENCY RELEASE

WARNING
The CPR emergency release is for emergency use only . When activating the CPR release handle, all people and
equipment must be removed from the area below and around the head, thigh and foot sections of the bed or serious
personal injury and/or equipment damage could occur .
When quick access to the patient is needed and the Fowler is raised, pull outward one of the two release handles
located under the upper right and left sides of the head section (see "Product Illustration" on
page
1-7) . The Fowler
will automatically flatten, and if raised, so will the Gatch .
The CPR handle can be released at any time to stop the lowering movement of the Fowler . But doing so will subsequently
require that the Fowler be completely lowered, using the CPR handles or the Fowler down control, to enable the Fowler
motor to reset itself . Failing to do so will prevent the Fowler from being fully raised .
Note
The use of the CPR release handle to partially lower the Fowler creates a situation where the course of the Fowler
motor is temporarily out of sync with the actual position of the Fowler . The situation is automatically corrected when the
Fowler is completely lowered . During the time the resetting process (approximately 30 seconds), the Fowler controls
are not available .
Lowering the Fowler During a Power Failure
The CPR emergency release can also be used during a power failure to partially or completely lower the Fowler . Simply
pull one of the CPR handles until the desired angle is reached . When power is restored, fully lower the Fowler to enable
the Fowler motor to reset itself .

NURSE CALL USAGE (OPTIONAL)

The nurse call function allows the patient to communicate through a speakerphone with the nurse station by simply
pressing the red nurse call
button on the inner control panel . The speakerphones are located on both head end
siderails (see "Product Illustration" on
page
1-7) .
The communication between the patient and the nurse station is established the moment a member of the nursing staff
responds to the call signal .
Note
The nurse call system will automatically send a call signal to the nurse station if the communication between the bed
and the nurse station is interrupted following a power failure, the switching off of the bed or the disconnection of the
nurse call communication cable .

AUXILIARY POWER OUTLET USAGE (OPTION AVAILABLE WITH 120VAC BED MODEL)

Located on the left side at the foot end of the bed (see "Product Illustration" on
page
1-7), this feature provides
nursing staff with a convenient power source for peripheral equipment . A 5A breaker is integrated to the power outlet .
WARNING
Use only hospital grade electric equipment consuming 5A or less with the auxiliary power outlet (optional) . The use of
standard electric equipment may bring the current leakage to an unacceptable level for hospital equipment .
Return To Table of Contents