Access Door Hinge Replacement - Stryker CUB Pediatric Crib Manuel De Maintenance

Masquer les pouces Voir aussi pour CUB Pediatric Crib:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

ACCESS DOOR HINGE REPLACEMENT

TOOLS REQUIRED:
• #1 Phillips Screwdriver
• Small Regular Screwdriver
• 5/32 in. Allen Wrench
• OG2 Grease
PROCEDURE:
Note
Unless otherwise indicated, refer to Figure 28,
1.
Remove the top half of the Siderail Upper Plastic Cover (see "Upper Plastic Cover Replacement",
2.
Remove the access door (see "Access Door
3.
Remove the colored label affixed to the outside of the lower cover of the access door.
4.
Using a #1 Phillips screwdriver, remove the four screws (O) holding the bottom half of the lower cover (P) to the
top half of the lower cover (Y).
5.
Insert a small regular screwdriver between the two halves of the lower cover to separate them. Use caution to
avoid scratching the covers or damaging the snap pins. Remove the bottom half of the access door lower cover.
6.
Using a 5/32 in. Allen wrench, remove the two Allen screws (AA) holding the top half of the lower hinge mechanism
(I) to the siderail lower structural member. Discard the defective halves of the lower hinge mechanism.
7.
Remove the colored label affixed to the outside of the upper cover of the access door.
8.
Using a #1 Phillips screwdriver, remove the four screws (M) holding the top half of the upper cover (N) to the
bottom half of the upper cover (W).
9.
Insert a small regular screwdriver between the two halves of the upper cover to separate them. Use caution to
avoid scratching the covers or damaging the snap pins. Remove the top half of the access door upper cover.
10. Using a 5/32 in. Allen wrench, remove the two Allen screws (U) holding the bottom half of the upper hinge
mechanism (G) to the siderail upper structural member.
11. Using a 5/32 in. Allen wrench, remove the two Allen screws (Q) holding the spacer (R) and the top half of the
upper hinge mechanism (H) to the siderail upper structural member (S). Set the spacer aside and discard the
upper hinge mechanism.
Note
Apply OG2 grease to the new hinges before installing them.
12. Reverse the above steps to install the new upper and lower hinges and reinstall the access door.
13. Verify proper operation of the access door before returning the stretcher to service.
Return To Table of Contents
1-68
Service
page
1-66, for this procedure.
Replacement",page
1-66).
1900-009-002 REV B
page
1-62).
www.stryker.com

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Fl19f/h

Table des Matières