Endrail Central Column Spring Replacement - Stryker CUB Pediatric Crib Manuel De Maintenance

Masquer les pouces Voir aussi pour CUB Pediatric Crib:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

ENDRAIL CENTRAL COLUMN SPRING REPLACEMENT

WARNING
Only service technicians from Stryker or service personnel trained by Stryker should perform this procedure. Failure to
comply can result in serious damage to the bed and/or severe injury to users or patients.
TOOLS REQUIRED:
• #1 Phillips Screwdriver
• #2 Phillips Screwdriver
• Small Regular Screwdriver
• 1/2 in. Socket
• 3/8 in. Drive Ratchet
• 3/8 in. Combination Wrench
• 3/32 in. Allen Wrench
• Medium Strength Thread Locker (or equivalent)
• Adjustable Supports (or equivalent)
PROCEDURE:
Note
When ordering a central column assembly, specify if the stretcher is equipped with the double safety lock. Proper
warning labels are affixed to assemblies with out the double safety lock.
1.
Remove the mattress, raise the rails to the highest position and apply the brakes.
2.
Using a #1 Phillips screwdriver, remove the eight screws (A) holding the bottom half of the lower plastic cover (B)
to the top half of the lower plastic cover (C).
3.
Insert a small regular screwdriver between the two halves of the lower plastic cover to separate them. Use caution
to avoid scratching the covers or damaging the upper cover snap pins.
4.
Using a #2 Phillips screwdriver, remove the two screws (D) holding the lower end of the central column to the lower
structural member (E).
5.
Using a #1 Phillips screwdriver, remove the eight screws (F) holding the top half of the upper plastic cover (G) to
the bottom half of the upper plastic cover (H).
6.
Insert a small regular screwdriver between the two halves of the upper plastic cover to separate them. Use caution
to avoid scratching the covers or damaging the lower cover snap pins.
7.
Lower the rail to the 14 in. locking position and place adjustable supports under the rail to support it when the
central column is removed.
8.
Using a cordless drill and a #2 Phillips bit, remove the 14 screws (J) holding the upper end of the posts and the
central column to the upper structural member.
9.
Using a 1/2 in. socket, remove the two bolts and two washers (L) holding the upper end of the two guide posts to
the structural member (K).
10. Remove the lower plastic cover.
11. Lift the central column to remove it. Use caution to avoid damaging the seal.
Note
The seal must be replaced if it is damaged during the procedure. Do not install a damaged seal.
12. Using a 3/32 in. Allen wrench, remove the set screw (A) holding the handle to the defective central column shaft.
Return To Table of Contents
1-54
Service
1900-009-002 REV B
www.stryker.com