7.2.2 Open the patient care kit. The patient care kit is intended for SINGLE PATIENT
USE ONLY. When using a GentleTouch
pillow at least 20 minutes prior to use, in order to allow for complete expansion
of the pillow.
7.3 5943AP Jackson Spinal Surgery Top with Advanced Control Pad System
The 5996 Advanced Control Pad System™ (ACP) with TEMPUR-MED
specifically for use with the Jackson Spinal Top. Utilizing the ACP System, automatic
massage action and periodic pressure point stimulation is provided to the patient at the
site of the hip and thigh pads. The ACP System consists of three components that work
together:
1. 5996-5 Variable cycle controller 100-120 VAC or 5996-6 Variable cycle controller-
230 VAC
2. 5996-200 Mizuho OSI TEMPUR-MED
3. 5996-12 ACP Quick connect tubing
The ACP System is an electronically controlled, pneumatically actuated system with a
rating of IPX-4 waterproof rating. It operates at either 100-120V-AC, 50/60hz or at
230V-AC, 50hz.
(NW0491) for complete set-up and use instructions.
7.4 Patient Support Pad Operation
The patient support pads attach to the Jackson Spinal Surgery Top using the lock lever
on the pad base. To attach, slip the pad base over the frame of the spinal top and
rotate the lever to lock as shown in Figures 33 and 34.
Clamp post
Jackson spinal
surgery top frame
Figure 33: Unlocked position
MIZUHOSI 2009
Refer to the 5996 Advanced Control Pad System user guide
Pad support
Rotate to
lock
(Support pad shown from below)
Pillow, it is recommended to open the
®
patient support pads
®
Lock lever
Figure 34: Locked position
28
is designed
®
Clamp post
Jackson spinal
surgery top frame
Lock lever
NW0504 Rev. D